MS02 Chlorine or Chlorine Dioxide Sensor Cube
Brand: Burkert Fluid Control SystemsThe device is selectable to measure chlorine or chlorine dioxide in water. It is used within the Online Analysis System Type 8905 by being plugged into a spare fluidic backplane slot. The sensor cube contains a high precision membrane covered amperiometric sensor, based on microelectromechanical systems technology (MEMS). The measurement signal shows the Cl₂ or ClO₂ content within the sample water.
The chlorine measurement reflects either the available chlorine HOCl or, if coupled with a MS01 pH sensor cube for pH compensation, the free chlorine. The electrical and fluidic connections are made via the connection panel of the system. The sensor cube is communicating via buS, so the recognition at the Online Analysis System is fully automatic. When plugging into a system you will find the sensor in the list of buS members for further customized adjustments.
Features- Fully compatible with buS systems and a wide range of further analysis sensor cubes
- Optional pH compensated chlorine measurement
- Modular sensor cube for hot swap (exchange during operation)
- Minimal sample water flow needed
- MEMS technology sensor
Specifications
Ambient Temperature Range
- Operating: 0° to 40° C (32° to 104° F)
- Storage: -10° to 60° C (14° to 140° F)
Approvals & Certifications
- CE
- EU Directives
- NAMUR NE 107 (External Communication by Status LED)
- UL (C,US)
Conductivity
- > 50 µS/cm
Consumption
- Power: 0.8 VA
Electrical Connectors
- Plug Connection
Enclosure, Body Material
- Polyphenylene Ether (PPE)
- Polystyrene (PS)
Environmental Protection
- IP20
- IP65
Flow Rate
- Sample Water: > 6 L/hour
Humidity
- < 90% Relative Humidity, without Condensation
Indication
- LED
Linearity
- ± 0.01 ppm or ± 3% of Measured Value, whichever is greater - ClO₂
- ± 0.02 ppm of Measured Value - Cl₂
Materials of Construction
- Lever: Polycarbonate (PC)
Measuring Error
- Measurement Deviation: ± 0.005 ppm or ± 3% of Measured Value, whichever is greater - ClO₂
- Measurement Deviation: ± 0.03 ppm or ± 5% of Measured Value - Cl₂
Measuring Range
- 0.005 to 5 ppm - ClO₂
- 0.01 to 5 ppm - Cl₂
Media
- Water (5 to 9 PH)
Operating Voltages
- 24 Vdc
Pressure Range
- Sample Water Pressure - psi: 0 to 43.51 (kPa: 0 to 300, bar: 0 to 3)
Process Temperature
- Sample Water Temperature: 3° to 40° C (37° to 104° F)
Repeatability
- ± 0.01 ppm or ± 3% of Measured Value, whichever is greater - ClO₂
- ± 0.02 ppm of Measured Value - Cl₂
Resolution
- Sensor: 0.001 ppm - ClO₂
- Sensor: 0.01 ppm - Cl₂
Response Time
- t90: < 30 seconds - Cl₂ & ClO₂
RTD Type
- Pt1000 Class B
Seat / Seal Material
- Seal: EPDM
Sensitivity
- -11 NA/ppm (at PH 5) - Cl₂
- -4 NA/ppm - ClO₂
- -8 NA/ppm (at PH 7) - Cl₂
Temperature Range
- 0° to 50° C (32° to 122° F)