Accuracy
- ± 0.5% of Full Scale
- ± 1% of Full Scale
Adjustment
- Zero: ± 33% of Full Scale
Applications
- Clean Room Monitoring
- Filter Monitoring
- Flow Measurement
- Fume Hoods
Approvals & Certifications
- CE EN 61000-4-2, EN 61000-4-3
- RoHS
Burst Pressure
- psi: 24.65 (kPa: 170, bar: 1.7)
Compensated Temperature
- 2° to 54° C (35° to 129° F)
Electrical Connectors
- 1/2" Female DIN Cable Conduit
- PG9 Cable Gland
- Screw Terminal
Humidity
- 10 to 90% Relative Humidity, Non-Condensing
Hysteresis
- ± 0.5% of Full Scale
- ± 1% of Full Scale
Linearity
- ± 0.5% of Full Scale
- ± 1% of Full Scale
Materials of Construction
- Cable Connection Socket: Nylon
- Polyester Foil (Front)
Media
- Clean, Dry Air
- Non-Corrosive Dry Gases
Mounting
- DIN Rail
- Panel Mount
- Wall Mounted
Operating Temperature
- -20° to 70° C (-4° to 158° F)
Proof Pressure
- psi: 14.5 (kPa: 100, bar: 1)
Repeatability
- ± 0.5% of Full Scale
- ± 1% of Full Scale
- ≤ ± 0.1% of Full Scale
Sensor, Probe Material
- Sensor: Aluminum
- Sensor: Glass
- Sensor: Silicon
Span & Zero Effect
- Adjustment: ± 5% of Full Scale
Stability
- ≤ 0.5% of Full Scale/Year
Storage Temperature
- -40° to 90° C (-40° to 194° F)
Supply Voltage
- 19-36 Vdc (Nominal 24 Vdc)
- Nominal 24 Vdc
Temperature Coefficient
- Zero & Span: ± 0.06%/K within 2° to 54° C (35° to 129° F), Reference Temperature 25° C (77° F)
Tube Connection
- 1/4" Barbed Tube Fitting Male